CYTOSPIN
Specification:
- The process at least 12 specimens at one time.
- Cyto-centrifuge lid is having one-handed opening and closing.
- The lid remains locked at all times during rotation.
- The operator is able to see the sealed head during cyto-centrifugation.
- Meet IEC 61010 standards for centrifuge safety.
- A pull-out program card allows for up to nine programs to be logged for fast and convenient retrieval.
The sealed head have the following features for Safety Propose:
- Auto-locking, plastic outer lid.
- Autoclavable sealed head.
- Disposable sample chambers with caps.
- Safety alarms that protect users and specimens.
- Wipe-clean control panel.
Controls:
- The Cyto-centrifuge have a protected program memory that stores up to 23 routines for instant recall and
protects from power loss.
- The unit able to spin at speeds from 200 – 2000 rpm.
- The unit has a minimum of three acceleration control buttons.
- Contain a window displays the programmed and remaining run time from 1– 99 minutes.
- Contain a standby power saving mode.
- Safety alarms that signals if the rotor is out of balance, outside the speed tolerance, or if the lid is not properly
locked.
- As an extra measure of safety for the user, the unit will not spin if the lid is not locked.